stoner_measurement.instruments.keithley.k2000.Keithley2000¶
- class stoner_measurement.instruments.keithley.k2000.Keithley2000(transport: BaseTransport, protocol: BaseProtocol | None = None)[source]¶
Bases:
DigitalMultimeterDriver for the Keithley 2000 digital multimeter.
- Variables:
transport (BaseTransport) – Transport layer (serial, GPIB, or Ethernet).
protocol (BaseProtocol) – Protocol instance (defaults to
ScpiProtocol).
- AC_FUNCTIONS = frozenset({DmmFunction.CURR_AC, DmmFunction.VOLT_AC})¶
- GATED_FUNCTIONS = frozenset({DmmFunction.FREQ, DmmFunction.PER})¶
- RANGED_FUNCTIONS = frozenset({DmmFunction.CURR_AC, DmmFunction.CURR_DC, DmmFunction.FRES, DmmFunction.RES, DmmFunction.VOLT_AC, DmmFunction.VOLT_DC})¶
- SUPPORTED_FUNCTIONS = (DmmFunction.VOLT_DC, DmmFunction.VOLT_AC, DmmFunction.CURR_DC, DmmFunction.CURR_AC, DmmFunction.RES, DmmFunction.FRES, DmmFunction.FREQ, DmmFunction.PER, DmmFunction.TEMP)¶
- configure_single_reading(function: DmmFunction, *, autorange: bool = True, input_range: float = 10.0, nplc: float = 1.0, filter_enabled: bool = False, filter_count: int = 10, filter_type: str = 'REP', trigger_delay: float = 0.0, ac_bandwidth: float = 300.0, gate_time: float = 0.1) None[source]¶
Configure one immediate DMM reading without operating scanner relays.
- Parameters:
function (DmmFunction) – Active measurement function.
- Keyword Parameters:
autorange (bool) – Select automatic ranging for ranged functions.
input_range (float) – Fixed input range in the function’s SI unit.
nplc (float) – Integration time in PLC, 0.01..10 where applicable.
filter_enabled (bool) – Enable digital averaging where supported.
filter_count (int) – Averaging count, 1..100.
filter_type (str) – REPeating (REP) or MOVing averaging.
trigger_delay (float) – Delay before conversion, in seconds.
ac_bandwidth (float) – AC lower bandwidth: 3, 30 or 300 Hz.
gate_time (float) – Frequency/period gate time, 0.01..1 seconds.
Temperature readings use kelvin and the instrument’s existing sensor configuration. The method sends no reset, route, scan-list or relay commands. No channel lists are used on the 2700.
- Raises:
ValueError – If the function or applicable settings are invalid.
- get_autorange() bool[source]¶
Return
Trueif autorange is enabled.- Returns:
Truewhen autorange is active.- Return type:
(bool)
- get_buffer_count() int[source]¶
Return the number of readings currently stored in the buffer.
- Returns:
Number of readings in the trace buffer.
- Return type:
(int)
- get_capabilities() DmmCapabilities[source]¶
Return static capability metadata for the Keithley 2000.
- Returns:
Capability descriptor.
- Return type:
- get_filter_count() int[source]¶
Return the averaging filter count.
- Returns:
Configured number of readings averaged per sample.
- Return type:
(int)
- get_filter_enabled() bool[source]¶
Return
Trueif digital averaging filter is enabled.- Returns:
Truewhen the averaging filter is active.- Return type:
(bool)
- get_measure_function() DmmFunction[source]¶
Return the active measurement function.
- Returns:
Active measurement function.
- Return type:
- get_nplc() float[source]¶
Return the integration time in power-line cycles.
- Returns:
Integration time in power-line cycles.
- Return type:
(float)
- get_range() float[source]¶
Return the active measurement range in units of the active function.
- Returns:
Active measurement range.
- Return type:
(float)
- get_trigger_count() int[source]¶
Return the configured trigger count.
- Returns:
Number of triggers configured.
- Return type:
(int)
- get_trigger_source() DmmTriggerSource[source]¶
Return the trigger source selection.
- Returns:
Active trigger source.
- Return type:
- measure() float[source]¶
Trigger a measurement and return its value.
- Returns:
Measured scalar value in units of the active function.
- Return type:
(float)
- read_buffer(count: int | None = None) tuple[float, ...][source]¶
Read values from the instrument trace buffer.
- Keyword Parameters:
count (int | None) – Number of points to read from the start of the buffer. If
None, read all available points.- Returns:
Parsed buffer readings.
- Return type:
- Raises:
ValueError – If count is not positive.
- read_single_snapshot(function: DmmFunction) tuple[float, dict[str, float]][source]¶
Read one fresh value and applicable numeric settings from the instrument.
- Parameters:
function (DmmFunction) – Expected configured function.
- Returns:
- Reading and queried settings. Filter
type is encoded as 0 for repeating and 1 for moving. Timeout is the actual transport timeout, not an instrument readback.
- Return type:
- Raises:
RuntimeError – If the function changed outside the plugin.
ValueError – If the response is empty, malformed or has multiple readings.
- set_autorange(state: bool) None[source]¶
Enable or disable autorange.
- Parameters:
state (bool) –
Trueto enable autorange.
- set_filter_count(count: int) None[source]¶
Set the averaging filter count.
- Parameters:
count (int) – Number of readings to average. Must be positive.
- Raises:
ValueError – If count is not positive.
- set_filter_enabled(state: bool) None[source]¶
Enable or disable the digital averaging filter.
- Parameters:
state (bool) –
Trueto enable filtering.
- set_measure_function(function: DmmFunction) None[source]¶
Set the active measurement function.
- Parameters:
function (DmmFunction) – Function to select.
- set_nplc(value: float) None[source]¶
Set the integration time in power-line cycles.
- Parameters:
value (float) – Integration time in power-line cycles. Must be positive.
- Raises:
ValueError – If value is not positive.
- set_range(value: float) None[source]¶
Set the measurement range.
- Parameters:
value (float) – Range value in units of the active function. Must be positive.
- Raises:
ValueError – If value is not positive.
- set_trigger_count(count: int) None[source]¶
Set the trigger count.
- Parameters:
count (int) – Number of triggers. Must be positive.
- Raises:
ValueError – If count is not positive.
- set_trigger_source(source: DmmTriggerSource) None[source]¶
Set the trigger source.
- Parameters:
source (DmmTriggerSource) – Trigger source to select.
- classmethod uses_nplc(function: DmmFunction, ac_bandwidth: float = 300.0) bool[source]¶
Return whether NPLC controls integration for this function and bandwidth.