stoner_measurement.instruments.lakeshore.ls625.Lakeshore625

class stoner_measurement.instruments.lakeshore.ls625.Lakeshore625(transport: BaseTransport, protocol: BaseProtocol | None = None)[source]

Bases: MagnetController, MagnetSupply

Driver for a Lakeshore 625 superconducting magnet power supply.

The driver composes BaseInstrument communication behaviour with the MagnetSupply interface for magnet supplies.

Variables:
  • transport (BaseTransport) – Transport used for instrument I/O.

  • protocol (BaseProtocol) – Protocol used to format and parse instrument messages.

Examples

>>> from stoner_measurement.instruments.lakeshore import Lakeshore625
>>> from stoner_measurement.instruments.transport import NullTransport
>>> t = NullTransport(responses=[b"LAKESHORE,MODEL625,SN001,1.0\r\n"])
>>> mps = Lakeshore625(transport=t)
>>> mps.connect()
>>> mps.identify()
'LAKESHORE,MODEL625,SN001,1.0'
>>> mps.disconnect()
abort_ramp() → None[source]

Stop ramping immediately.

The Model 625 command set does not provide a dedicated abort command, so this maps to the supported STOP action.

property current: float

Return output current in amps.

Returns:

Measured magnet supply current in amps.

Return type:

(float)

property field: float

Return output field in tesla.

Returns:

Measured magnetic field in tesla.

Return type:

(float)

get_firmware_version() → str[source]

Return the firmware version from the identification string.

Returns:

Firmware version token when available, otherwise an empty string.

Return type:

(str)

get_model() → str[source]

Return the model name from the identification string.

Returns:

Instrument model token when available, otherwise an empty string.

Return type:

(str)

go_to_zero() → None[source]

Ramp the supply output to zero.

property heater: bool

Return persistent switch heater state.

Returns:

True when the heater is stably on.

Return type:

(bool)

heater_off() → None[source]

Disable the persistent switch heater.

heater_on() → None[source]

Enable the persistent switch heater.

hold() → None[source]

Hold the present output without changing field.

property limits: MagnetLimits

Return output limits read from the instrument.

Returns:

Current, derived field, and current-ramp limits. The Model 625 reports ramp-rate limits in A/s, which are converted to T/min.

Return type:

(MagnetLimits)

property magnet_constant: float

Return the cached instrument field constant in tesla per amp.

Returns:

Magnet constant in tesla per amp.

Return type:

(float)

pause_ramp() → None[source]

Pause an active ramp.

property ramp_rate_current: float | None

Return the programmed current ramp rate in amps per minute.

property ramp_rate_field: float | None

Return the programmed field ramp rate in tesla per minute.

ramp_to_current(current: float, *, wait: bool = False) → None[source]

Program current target and ramp.

Parameters:

current (float) – Target current in amps.

Keyword Parameters:

wait (bool) – When True, block until ramp completes.

ramp_to_field(field: float, *, wait: bool = False) → None[source]

Program field target and ramp.

Parameters:

field (float) – Target field in tesla.

Keyword Parameters:

wait (bool) – When True, block until ramp completes.

ramp_to_target() → None[source]

Start ramping to the current programmed target.

refresh_magnet_constant() → float[source]

Query and cache the instrument field constant in tesla per amp.

return_to_local() → None[source]

No-op for the Lakeshore 625, which has no Oxford-style local handoff command.

set_limits(limits: MagnetLimits) → None[source]

Set software limits used by higher-level sequence logic.

Parameters:

limits (MagnetLimits) – Limit configuration to cache for runtime checks.

set_magnet_constant(tesla_per_amp: float) → None[source]

Set the software magnet constant used for conversion.

Parameters:

tesla_per_amp (float) – Magnet constant in tesla per amp.

set_ramp_rate_current(rate: float) → None[source]

Set the current ramp rate in amps per minute.

Parameters:

rate (float) – Ramp rate in amps per minute.

set_ramp_rate_field(rate: float) → None[source]

Set the field ramp rate in tesla per minute.

Parameters:

rate (float) – Ramp rate in tesla per minute.

set_target_current(current: float) → None[source]

Set the target current in amps.

Parameters:

current (float) – Target current in amps.

set_target_field(field: float) → None[source]

Set the target field in tesla.

Parameters:

field (float) – Target magnetic field in tesla.

property status: MagnetStatus

Return consolidated magnet status.

Queries OPST? which returns a numeric bit-coded operating status.

Returns:

Snapshot of controller state and key readings.

Return type:

(MagnetStatus)

property target_current: float | None

Return the programmed current target in amps.

property target_field: float | None

Return the programmed field target in tesla.

property voltage: float

Return output voltage in volts.

Returns:

Measured output voltage in volts.

Return type:

(float)