stoner_measurement.instruments.lockin_amplifier.LockInAmplifier

class stoner_measurement.instruments.lockin_amplifier.LockInAmplifier(transport: BaseTransport, protocol: BaseProtocol)[source]

Bases: BaseInstrument

Abstract base class for lock-in amplifier instruments.

Provides a common interface for dual-output lock-in amplifiers where the measured signal is available as in-phase/quadrature components (X, Y) and polar components (magnitude R, angle theta).

Variables:
auto_gain() → None[source]

Run instrument auto-gain adjustment.

Raises:

NotImplementedError – If auto-gain is not supported by the instrument.

auto_phase() → None[source]

Run instrument auto-phase adjustment.

Raises:

NotImplementedError – If auto-phase is not supported by the instrument.

auto_reserve() → None[source]

Run instrument auto-reserve adjustment.

Raises:

NotImplementedError – If auto-reserve is not supported by the instrument.

abstractmethod get_capabilities() → LockInAmplifierCapabilities[source]

Return static capability metadata.

Returns:

Capability descriptor.

Return type:

(LockInAmplifierCapabilities)

get_dynamic_reserve_db() → float[source]

Return the dynamic reserve (signal stability) in decibels.

A higher value indicates that the instrument can tolerate larger interfering signals relative to the signal of interest without introducing significant errors.

Returns:

Dynamic reserve in dB.

Return type:

(float)

Raises:

NotImplementedError – If numeric dynamic reserve control is not supported by the instrument.

get_filter_slope() → int[source]

Return the output filter roll-off slope in dB/octave.

Raises:

NotImplementedError – If filter-slope control is not supported by the instrument.

get_harmonic() → int[source]

Return the selected detection harmonic.

Raises:

NotImplementedError – If harmonic selection is not supported by the instrument.

get_input_coupling() → LockInInputCoupling[source]

Return the input coupling mode.

Raises:

NotImplementedError – If input-coupling control is not supported by the instrument.

get_input_shielding() → LockInInputShielding[source]

Return the input shield grounding mode.

Returns:

Active input shield grounding mode.

Return type:

(LockInInputShielding)

Raises:

NotImplementedError – If input shielding control is not supported by the instrument.

get_input_source() → LockInInputSource[source]

Return the active input source.

Returns:

Active input source selection.

Return type:

(LockInInputSource)

Raises:

NotImplementedError – If input source selection is not supported by the instrument.

get_line_filter() → LockInLineFilter[source]

Return the line-frequency notch filter configuration.

Returns:

Active notch filter configuration.

Return type:

(LockInLineFilter)

Raises:

NotImplementedError – If line filter control is not supported by the instrument.

get_oscillator_amplitude() → float[source]

Return the internal oscillator sine output amplitude in volts.

Returns:

Oscillator amplitude in volts.

Return type:

(float)

Raises:

NotImplementedError – If the instrument does not have a controllable internal oscillator.

get_output_offset(channel: LockInOutputChannel) → tuple[float, LockInExpandFactor][source]

Return the output offset percentage and expand factor for a channel.

Parameters:

channel (LockInOutputChannel) – Output channel to query.

Returns:

(offset_pct, expand_factor) where offset_pct is the offset as a percentage and expand_factor is the expand multiplier.

Return type:

(tuple[float, LockInExpandFactor])

Raises:

NotImplementedError – If output offset and expand are not supported by the instrument.

abstractmethod get_reference_frequency() → float[source]

Return the reference frequency in hertz.

Returns:

Reference frequency in hertz.

Return type:

(float)

abstractmethod get_reference_phase() → float[source]

Return the reference phase in degrees.

Returns:

Reference phase in degrees.

Return type:

(float)

abstractmethod get_reference_source() → tuple[LockInReferenceSource, LockinRefenceEdge][source]

Return the active reference source.

Returns:

Active reference source.

Return type:

(LockInReferenceSource)

get_reserve_mode() → LockInReserveMode[source]

Return the dynamic reserve mode.

Raises:

NotImplementedError – If reserve-mode control is not supported by the instrument.

abstractmethod get_sensitivity() → float[source]

Return the active sensitivity scale.

Returns:

Sensitivity in volts.

Return type:

(float)

get_sync_filter_enabled() → bool[source]

Return whether the synchronous output filter is enabled.

Returns:

True when the synchronous filter is enabled.

Return type:

(bool)

Raises:

NotImplementedError – If synchronous filter control is not supported by the instrument.

abstractmethod get_time_constant() → float[source]

Return the active output filter time constant.

Returns:

Time constant in seconds.

Return type:

(float)

abstractmethod measure_rt() → tuple[float, float][source]

Measure and return magnitude and phase outputs.

Returns:

(magnitude, theta) where magnitude is in volts and theta is in degrees.

Return type:

(tuple[float, float])

abstractmethod measure_xy() → tuple[float, float][source]

Measure and return the in-phase and quadrature outputs.

Returns:

(x, y) values in volts, where x is the in-phase component and y is the quadrature component.

Return type:

(tuple[float, float])

set_dynamic_reserve_db(value_db: float) → None[source]

Set the dynamic reserve (signal stability) in decibels.

Parameters:

value_db (float) – Desired dynamic reserve in dB.

Raises:

NotImplementedError – If numeric dynamic reserve control is not supported by the instrument.

set_filter_slope(slope: int) → None[source]

Set the output filter roll-off slope in dB/octave.

Raises:

NotImplementedError – If filter-slope control is not supported by the instrument.

set_harmonic(harmonic: int) → None[source]

Set the selected detection harmonic.

Raises:

NotImplementedError – If harmonic selection is not supported by the instrument.

set_input_coupling(coupling: LockInInputCoupling) → None[source]

Set the input coupling mode.

Raises:

NotImplementedError – If input-coupling control is not supported by the instrument.

set_input_shielding(shielding: LockInInputShielding) → None[source]

Set the input shield grounding mode.

Parameters:

shielding (LockInInputShielding) – Shielding mode to select.

Raises:

NotImplementedError – If input shielding control is not supported by the instrument.

set_input_source(source: LockInInputSource) → None[source]

Set the active input source.

Parameters:

source (LockInInputSource) – Input source to select.

Raises:

NotImplementedError – If input source selection is not supported by the instrument.

set_line_filter(filter_config: LockInLineFilter) → None[source]

Set the line-frequency notch filter configuration.

Parameters:

filter_config (LockInLineFilter) – Notch filter configuration to apply.

Raises:

NotImplementedError – If line filter control is not supported by the instrument.

set_oscillator_amplitude(value: float) → None[source]

Set the internal oscillator sine output amplitude in volts.

Parameters:

value (float) – Amplitude in volts.

Raises:

NotImplementedError – If the instrument does not have a controllable internal oscillator.

set_output_offset(channel: LockInOutputChannel, offset_pct: float, expand_factor: LockInExpandFactor) → None[source]

Set the output offset and expand factor for a channel.

Parameters:
Raises:

NotImplementedError – If output offset and expand are not supported by the instrument.

abstractmethod set_reference_frequency(value: float) → None[source]

Set the reference frequency in hertz.

Parameters:

value (float) – Frequency in hertz.

abstractmethod set_reference_phase(value: float) → None[source]

Set the reference phase in degrees.

Parameters:

value (float) – Phase in degrees.

abstractmethod set_reference_source(source: LockInReferenceSource, edge: LockinRefenceEdge = LockinRefenceEdge.FALLING) → None[source]

Set the active reference source.

Parameters:

source (LockInReferenceSource) – Source to select.

set_reserve_mode(mode: LockInReserveMode) → None[source]

Set the dynamic reserve mode.

Raises:

NotImplementedError – If reserve-mode control is not supported by the instrument.

abstractmethod set_sensitivity(value: float) → None[source]

Set the active sensitivity scale.

Parameters:

value (float) – Sensitivity in volts.

set_sync_filter_enabled(state: bool) → None[source]

Enable or disable the synchronous output filter.

Parameters:

state (bool) – True to enable the synchronous filter.

Raises:

NotImplementedError – If synchronous filter control is not supported by the instrument.

abstractmethod set_time_constant(value: float) → None[source]

Set the output filter time constant.

Parameters:

value (float) – Time constant in seconds.