stoner_measurement.instruments.nanovoltmeter.Nanovoltmeter¶
- class stoner_measurement.instruments.nanovoltmeter.Nanovoltmeter(transport: BaseTransport, protocol: BaseProtocol)[source]¶
Bases:
BaseInstrumentAbstract base class for nanovoltmeter instruments.
Provides a uniform interface for high-sensitivity DC voltage measurements. Nanovoltmeters are commonly used in low-resistance measurements, Hall-effect measurements, and thermoelectric studies.
- Variables:
transport (BaseTransport) – Transport layer instance.
protocol (BaseProtocol) – Protocol instance.
Examples
>>> # Demonstrate interface using a minimal concrete implementation >>> from stoner_measurement.instruments.transport import NullTransport >>> from stoner_measurement.instruments.protocol import ScpiProtocol >>> from stoner_measurement.instruments.nanovoltmeter import Nanovoltmeter >>> class _NVM(Nanovoltmeter): ... def measure_voltage(self): return 1.23e-6 ... def get_range(self): return 0.1 ... def set_range(self, value): pass ... def get_autorange(self): return True ... def set_autorange(self, state): pass ... def get_nplc(self): return 5.0 ... def set_nplc(self, value): pass >>> nvm = _NVM(NullTransport(), ScpiProtocol()) >>> nvm.measure_voltage() 1.23e-06
- CAPABILITIES: ClassVar[NanovoltmeterCapabilities] = NanovoltmeterCapabilities(has_function_selection=False, has_filter=False, has_trigger=False, has_buffer=False, has_data_format=False, has_byte_order=False, supported_functions=(<NanovoltmeterFunction.VOLT: 'VOLT'>,), fixed_voltage_ranges=(), nplc_values=(), digit_values=(), filter_types=(), default_nplc=None, default_digits=None, default_filter_type=None, counted_filter_types=(), analog_filter_time_multiplier=1.0, supports_filter_count=False, supports_analog_filter=False, supports_relative=False, supports_line_sync=False, supports_autozero=False, supports_safe_reset=True, relative_limits=None, max_buffer_points=None)¶
- abort() None[source]¶
Abort a running measurement sequence.
- Raises:
NotImplementedError – If trigger configuration is not supported by the instrument.
- clear_buffer() None[source]¶
Clear the instrument reading buffer.
- Raises:
NotImplementedError – If buffer operations are not supported by the instrument.
- abstractmethod get_autorange() bool[source]¶
Return
Trueif autorange is currently enabled.- Returns:
Truewhen the instrument selects the range automatically.- Return type:
(bool)
- Raises:
ConnectionError – If the transport is not open.
- get_buffer_count() int[source]¶
Return the number of readings currently stored in the buffer.
- Raises:
NotImplementedError – If buffer operations are not supported by the instrument.
- abstractmethod get_capabilities() NanovoltmeterCapabilities[source]¶
Return static capability metadata.
- Returns:
Capability descriptor.
- Return type:
- get_filter_count() int[source]¶
Return the configured filter averaging count.
- Raises:
NotImplementedError – If filtering is not supported by the instrument.
- get_filter_enabled() bool[source]¶
Return whether filtering is enabled.
- Raises:
NotImplementedError – If filtering is not supported by the instrument.
- abstractmethod get_measure_function() NanovoltmeterFunction[source]¶
Return the active measurement function.
- Returns:
Active measurement function.
- Return type:
- abstractmethod get_nplc() float[source]¶
Return the integration time in power-line cycles.
- Returns:
Integration time in power-line cycles.
- Return type:
(float)
- Raises:
ConnectionError – If the transport is not open.
- abstractmethod get_range() float[source]¶
Return the current voltage measurement range in volts.
- Returns:
Active measurement range in volts.
0.0typically indicates autorange is active.- Return type:
(float)
- Raises:
ConnectionError – If the transport is not open.
- get_trigger_count() int[source]¶
Return the configured trigger count.
- Raises:
NotImplementedError – If trigger configuration is not supported by the instrument.
- get_trigger_source() NanovoltmeterTriggerSource[source]¶
Return the trigger source selection.
- Raises:
NotImplementedError – If trigger configuration is not supported by the instrument.
- initiate() None[source]¶
Arm the trigger system and begin measurements.
- Raises:
NotImplementedError – If trigger configuration is not supported by the instrument.
- abstractmethod measure_voltage() float[source]¶
Trigger a voltage measurement and return the result in volts.
- Returns:
Measured voltage in volts.
- Return type:
(float)
- Raises:
ConnectionError – If the transport is not open.
- read_buffer(count: int | None = None) tuple[float, ...] | ndarray[source]¶
Read values from the instrument buffer.
- Keyword Parameters:
count (int | None) – Optional number of points to read from the start of the buffer. If
None, read all available points.- Returns:
Parsed buffer values.
- Return type:
- Raises:
NotImplementedError – If buffer operations are not supported by the instrument.
- set_analog_filter_enabled(state: bool) None[source]¶
Enable or disable analogue low-pass filtering.
- Raises:
NotImplementedError – If analogue filtering is not supported by the instrument.
- abstractmethod set_autorange(state: bool) None[source]¶
Enable or disable automatic range selection.
- Parameters:
state (bool) –
Trueto enable autorange,Falseto use the manually set range.- Raises:
ConnectionError – If the transport is not open.
- set_buffer_feed_continuous_next() None[source]¶
Set buffer feed mode to continuous-next.
- Raises:
NotImplementedError – If configurable buffer feed mode is not supported.
- set_buffer_feed_sense() None[source]¶
Set buffer feed source to measurement readings.
- Raises:
NotImplementedError – If configurable buffer feed source is not supported.
- set_buffer_size(size: int) None[source]¶
Set reading-buffer capacity.
- Raises:
NotImplementedError – If buffer sizing is not supported by the instrument.
- set_digits(digits: int) None[source]¶
Set display/data resolution digits.
- Raises:
NotImplementedError – If digit-resolution control is not supported by the instrument.
- set_filter_count(count: int) None[source]¶
Set the filter averaging count.
- Raises:
NotImplementedError – If filtering is not supported by the instrument.
- set_filter_enabled(state: bool) None[source]¶
Enable or disable measurement filtering.
- Raises:
NotImplementedError – If filtering is not supported by the instrument.
- abstractmethod set_measure_function(function: NanovoltmeterFunction) None[source]¶
Set the active measurement function.
- Parameters:
function (NanovoltmeterFunction) – Function to select.
- abstractmethod set_nplc(value: float) None[source]¶
Set the integration time in power-line cycles.
- Parameters:
value (float) – Integration time in power-line cycles. Longer integration times improve resolution but reduce throughput.
- Raises:
ConnectionError – If the transport is not open.
ValueError – If value is outside the valid range for this instrument.
- abstractmethod set_range(value: float) None[source]¶
Set the voltage measurement range.
- Parameters:
value (float) – Measurement range in volts. Pass
0.0to enable autorange on instruments that conflate the two settings.- Raises:
ConnectionError – If the transport is not open.
ValueError – If value is not a valid range for this instrument.
- set_relative_enabled(state: bool) None[source]¶
Enable or disable relative (REL) mode.
- Raises:
NotImplementedError – If relative mode is not supported by the instrument.
- set_trigger_count(count: int) None[source]¶
Set the trigger count.
- Raises:
NotImplementedError – If trigger configuration is not supported by the instrument.
- set_trigger_source(source: NanovoltmeterTriggerSource) None[source]¶
Set the trigger source selection.
- Raises:
NotImplementedError – If trigger configuration is not supported by the instrument.