stoner_measurement.instruments.oxford.mercury_ips.OxfordMercuryIPS

class stoner_measurement.instruments.oxford.mercury_ips.OxfordMercuryIPS(transport: BaseTransport, protocol: BaseProtocol | None = None, *, device_uid: str = 'GRPZ')[source]

Bases: MagnetController, MagnetSupply

Driver for an Oxford Instruments Mercury iPS magnet power supply.

The Mercury iPS uses a native text protocol, including over GPIB. Commands and queries follow the pattern:

READ:DEV:<uid>:PSU:SIG:<signal>
SET:DEV:<uid>:PSU:SIG:<signal>:<value>

Responses carry a STAT:DEV:<uid>:PSU:… prefix with the value appended after the final colon (e.g. STAT:DEV:PSU.M1:PSU:SIG:FLD:+1.23456T).

All field values are in tesla and ramp rates in tesla per minute.

Variables:
  • transport (BaseTransport) – Transport used for instrument I/O.

  • protocol (BaseProtocol) – Protocol used to format and parse instrument messages.

  • device_uid (str) – Mercury iPS device UID used in all READ/SET commands. Defaults to the single-axis group "GRPZ". Use the appropriate group UID for another axis; ACTN, STAT and SWHT are group settings.

Examples

>>> from stoner_measurement.instruments.oxford import OxfordMercuryIPS
>>> from stoner_measurement.instruments.transport import NullTransport
>>> t = NullTransport(
...     responses=[b"IDN:OXFORD INSTRUMENTS:MERCURY iPS:12345:2.7.0\n"] * 2,
... )
>>> mps = OxfordMercuryIPS(transport=t)
>>> mps.connect()
>>> mps.identify()
'IDN:OXFORD INSTRUMENTS:MERCURY iPS:12345:2.7.0'
>>> mps.disconnect()
abort_ramp() → None[source]

Abort an active ramp, holding the output at its current value.

Sends SET:DEV:<uid>:PSU:ACTN:HOLD to immediately stop any active ramp. To ramp to zero instead, use ramp_to_field() with a target of 0.0.

Raises:

ConnectionError – If the transport is not open.

property current: float

Return output current in amps.

Returns:

Measured magnet supply current in amps.

Return type:

(float)

Raises:

ConnectionError – If the transport is not open.

disconnect() → None[source]

Close the owned connection without changing the magnet output.

property field: float

Return output field in tesla.

Returns:

Measured magnetic field in tesla as reported by the instrument.

Return type:

(float)

Raises:

ConnectionError – If the transport is not open.

get_firmware_version() → str[source]

Return the firmware version from the identity string.

Returns:

Firmware field of the *IDN? response (e.g. "2.7.0"), or an empty string if not parseable.

Return type:

(str)

Raises:

ConnectionError – If the transport is not open.

Examples

>>> mps.get_firmware_version()
'2.7.0'
get_model() → str[source]

Return the model name from the identity string.

Returns:

Model field of the native or comma-separated *IDN? response, or an empty string if not parseable.

Return type:

(str)

Raises:

ConnectionError – If the transport is not open.

Examples

>>> mps.get_model()
'Mercury iPS'
go_to_zero() → None[source]

Ramp the supply output to zero.

Sends SET:DEV:<uid>:PSU:ACTN:RTOZ to the instrument.

property heater: bool

Return the persistent switch heater state.

Returns:

True when the switch heater is on, False when it is off.

Return type:

(bool)

Raises:

ConnectionError – If the transport is not open.

heater_off() → None[source]

De-energise the persistent switch heater.

Sends SET:DEV:<uid>:PSU:SIG:SWHT:OFF to the instrument.

Raises:

ConnectionError – If the transport is not open.

heater_on() → None[source]

Energise the persistent switch heater.

Sends SET:DEV:<uid>:PSU:SIG:SWHT:ON to the instrument.

Raises:

ConnectionError – If the transport is not open.

hold() → None[source]

Hold the present output without changing field.

Sends SET:DEV:<uid>:PSU:ACTN:HOLD to the instrument.

identify() → str[source]

Return the instrument identity string.

Returns:

Native identity string returned by *IDN?, normally IDN:OXFORD INSTRUMENTS:MERCURY iPS:<serial>:<firmware>.

Return type:

(str)

Raises:

ConnectionError – If the transport is not open.

Examples

>>> mps.identify()
'Oxford Instruments,Mercury iPS,12345,2.7.0'
property limits: MagnetLimits

Return the configured software operating limits.

Returns:

Cached current/field/ramp limits set via set_limits().

Return type:

(MagnetLimits)

property magnet_constant: float

Return the software magnet constant in tesla per amp.

Returns:

Field-to-current conversion factor in T A⁻¹ held by this driver instance.

Return type:

(float)

pause_ramp() → None[source]

Pause an active ramp, holding the output at its current value.

Sends SET:DEV:<uid>:PSU:ACTN:HOLD to the instrument.

Raises:

ConnectionError – If the transport is not open.

property ramp_rate_current: float | None

Return the programmed current ramp rate in amps per minute.

property ramp_rate_field: float | None

Return the programmed field ramp rate in tesla per minute.

ramp_to_current(current: float, *, wait: bool = False) → None[source]

Programme a current target and begin ramping.

Parameters:

current (float) – Target current in amps.

Keyword Parameters:

wait (bool) – When True, block until the ramp completes. Defaults to False.

Raises:

ConnectionError – If the transport is not open.

ramp_to_field(field: float, *, wait: bool = False) → None[source]

Programme a field target and begin ramping.

Parameters:

field (float) – Target field in tesla.

Keyword Parameters:

wait (bool) – When True, block until the ramp completes. Defaults to False.

Raises:

ConnectionError – If the transport is not open.

ramp_to_target() → None[source]

Start ramping to the currently programmed field setpoint.

Sends SET:DEV:<uid>:PSU:ACTN:RTOS to the instrument.

Raises:

ConnectionError – If the transport is not open.

return_to_local() → None[source]

Leave local/remote selection to the Mercury front panel.

The native protocol has no PSU:CLOC command. The driver never locks the front panel; select Local there as described in manual section 9.5.

set_limits(limits: MagnetLimits) → None[source]

Set software operating limits for this driver instance.

Parameters:

limits (MagnetLimits) – Limit configuration to cache for runtime checks.

set_magnet_constant(tesla_per_amp: float) → None[source]

Set the software magnet constant used for current/field conversions.

Parameters:

tesla_per_amp (float) – Field-to-current conversion factor in T A⁻¹.

Raises:

ValueError – If tesla_per_amp is not positive.

set_ramp_rate_current(rate: float) → None[source]

Set the current ramp rate.

Converts rate in amps per minute to tesla per minute using the configured magnet constant, then calls set_ramp_rate_field().

Parameters:

rate (float) – Ramp rate in amps per minute.

Raises:
set_ramp_rate_field(rate: float) → None[source]

Set the field ramp rate.

Parameters:

rate (float) – Ramp rate in tesla per minute.

Raises:
set_target_current(current: float) → None[source]

Set the target output current.

Converts current to a field setpoint using the configured magnet constant and sends a SET:DEV:<uid>:PSU:SIG:FSET command.

Parameters:

current (float) – Desired target current in amps.

Raises:

ConnectionError – If the transport is not open.

set_target_field(field: float) → None[source]

Set the target magnetic field.

Parameters:

field (float) – Desired target field in tesla.

Raises:

ConnectionError – If the transport is not open.

property status: MagnetStatus

Return one magnet snapshot under the shared instrument lock.

property target_current: float | None

Return the programmed current target in amps.

property target_field: float | None

Return the programmed field target in tesla.

property voltage: float

Return output voltage in volts.

Returns:

Measured output voltage in volts.

Return type:

(float)

Raises:

ConnectionError – If the transport is not open.

write(command: str, slow: int | None = None) → None[source]

Send a native SET and consume its matching acknowledgement.

Parameters:

command (str) – Complete native SET command without terminator.

Keyword Parameters:

slow (int | None) – Optional transport delay in milliseconds.

Raises: