stoner_measurement.instruments.temperature_controller.ControllerCapabilities¶
- class stoner_measurement.instruments.temperature_controller.ControllerCapabilities(num_inputs: int, num_loops: int, input_channels: tuple[str, ...], loop_numbers: tuple[int, ...], has_ramp: bool = True, has_pid: bool = True, has_autotune: bool = False, has_alarm: bool = False, has_zone: bool = False, has_user_curves: bool = False, has_sensor_excitation: bool = False, has_cryogen_control: bool = False, has_gas_auto_mode: bool = False, has_manual_heater_output: bool = False, has_input_settings: bool = False, heater_range_labels: dict[int, tuple[str, ...]] = <factory>, min_temperature: float | None = None, max_temperature: float | None = None, has_heater_range: bool = True, control_modes: tuple[~stoner_measurement.instruments.temperature_controller.ControlMode, ...] = (), heater_range_loops: tuple[int, ...] | None = None, setpoint_after_configuration: bool = False, input_settings_model: str = 'standard')[source]¶
Bases:
objectStatic capability descriptor for a temperature controller driver.
- Variables:
num_inputs (int) – Total number of sensor input channels.
num_loops (int) – Total number of PID control loops.
input_channels (tuple[str, ...]) – Ordered tuple of channel identifiers (e.g.
("A", "B", "C")).loop_numbers (tuple[int, ...]) – Ordered tuple of loop numbers (e.g.
(1, 2)).has_ramp (bool) –
Trueif the driver supports setpoint ramping.has_pid (bool) –
Trueif the driver supports PID parameter read/write.has_autotune (bool) –
Trueif the driver supports PID autotuning.has_alarm (bool) –
Trueif the driver supports sensor alarm limits.has_zone (bool) –
Trueif the driver supports zone/table (scheduled PID) control.has_user_curves (bool) –
Trueif the driver supports user-defined calibration curves.has_sensor_excitation (bool) –
Trueif the driver supports configuring sensor excitation.has_cryogen_control (bool) –
Trueif the driver supports cryogen-flow or needle-valve control (e.g. Oxford Instruments ITC503 or Mercury iTC).has_gas_auto_mode (bool) –
Trueif the driver supports switching the gas/needle-valve control between automatic and manual mode.has_manual_heater_output (bool) –
Trueif the driver supports setting the heater output directly (required forOPEN_LOOPoperation).has_heater_range (bool) –
Trueif the controller provides selectable heater power ranges.control_modes (tuple[ControlMode, ...]) – Supported modes, or an empty tuple to use the generic mode choices.
heater_range_loops (tuple[int, ...] | None) – Loops with discrete heater ranges. None means all loops when has_heater_range is true; an empty tuple means none.
setpoint_after_configuration (bool) – Configure ramp and output settings before writing the target in Apply.
input_settings_model (str) – Input editor schema. The default uses the shared 335/336 layout;
lakeshore340selects the Model 340 native input editor anditc503selects per-sensor software temperature scales.has_input_settings (bool) –
Trueif the driver supports reading and writing per-channel input configuration (sensor type, range, filter, calibration curve assignment) viaget_input_channel_settings()andset_input_channel_settings(), or through the specialised software-scale API foritc503.heater_range_labels (dict[int, tuple[str, ...]]) – Optional per-loop mapping of heater range index to human-readable label. The tuple index corresponds to the integer range index passed to
set_heater_range(). An empty dict (the default) means the UI should fall back to numeric indices.min_temperature (float | None) – Minimum achievable temperature in Kelvin, or
Noneif unknown.max_temperature (float | None) – Maximum controllable temperature in Kelvin, or
Noneif unknown.
- control_modes: tuple[ControlMode, ...] = ()¶