stoner_measurement.instruments.temperature_controller.ZoneEntry¶
- class stoner_measurement.instruments.temperature_controller.ZoneEntry(upper_bound: float, p: float, i: float, d: float, ramp_rate: float, heater_range: int, heater_output: float, input_channel: int = 0)[source]¶
Bases:
objectOne entry in a temperature controller’s zone / table control table.
Zone control divides the operating temperature range into consecutive segments. Each segment has its own PID gains, setpoint ramp rate, heater range, and manual heater output power. When the active setpoint crosses upper_bound the controller automatically loads the next zone’s parameters.
- Variables:
upper_bound (float) – Upper setpoint boundary for this zone in Kelvin. When the setpoint is at or below this value the zone’s PID and heater settings are applied.
p (float) – Proportional gain for this zone.
i (float) – Integral gain (reset) for this zone.
d (float) – Derivative gain (rate) for this zone.
ramp_rate (float) – Setpoint ramp rate for this zone in Kelvin per minute. A value of
0means the setpoint changes immediately (no ramp).heater_range (int) – Heater range index for this zone (instrument-specific;
0conventionally means heater off).heater_output (float) – Manual heater output power percentage for this zone (0–100 %). Used when the loop is in open-loop mode or as the initial value when entering the zone.
input_channel (int) – Instrument-specific zone control-input index.
0selects the loop’s default control input on Lakeshore controllers.