"""Single-reading command for Keithley 2000 and 2700 multimeters."""
from __future__ import annotations
import math
from typing import Any
from qtpy.QtWidgets import QCheckBox, QComboBox, QFormLayout, QSpinBox, QVBoxLayout, QWidget
from stoner_measurement.core.plugin_outputs import PluginOutputs, ScalarOutput
from stoner_measurement.core.validation import ValidationError, validate_expression
from stoner_measurement.instruments.dmm import DmmFunction
from stoner_measurement.instruments.keithley import Keithley2000, Keithley2700
from stoner_measurement.instruments.transport import GpibTransport
from stoner_measurement.plugins.command.base import CommandPlugin
from stoner_measurement.ui.widgets import FILTER_GPIB, SISpinBox, VisaResourceComboBox
DMM_DRIVERS = {"keithley_2000": Keithley2000, "keithley_2700": Keithley2700}
DMM_CHANNELS = {
DmmFunction.VOLT_DC: ("DC Voltage", "V"),
DmmFunction.VOLT_AC: ("AC Voltage", "V"),
DmmFunction.CURR_DC: ("DC Current", "A"),
DmmFunction.CURR_AC: ("AC Current", "A"),
DmmFunction.RES: ("Resistance", "Ω"),
DmmFunction.FRES: ("Four-wire Resistance", "Ω"),
DmmFunction.FREQ: ("Frequency", "Hz"),
DmmFunction.PER: ("Period", "s"),
DmmFunction.TEMP: ("Temperature", "K"),
}
_NUMERIC_SETTINGS = {
"input_range": (1e-12, 1e12),
"nplc": (0.01, 10.0),
"gate_time": (0.01, 1.0),
"trigger_delay": (0.0, 999999.999),
"timeout": (0.01, 86400.0),
}
_SAVED_SETTINGS = (
"driver",
"resource",
"function",
"autorange",
"input_range",
"nplc",
"filter_enabled",
"filter_count",
"filter_type",
"ac_bandwidth",
"gate_time",
"trigger_delay",
"timeout",
)
[docs]
class ReadMultimeterCommand(CommandPlugin):
"""Take one fresh reading from a Keithley 2000 or 2700 multimeter.
Add **Keithleys → Read Multimeter** beneath a scan or after a set command.
**General** selects the model, GPIB resource and measurement function: DC
or AC voltage/current, two- or four-wire resistance, frequency, period or
temperature. The selected reading is a scalar output with its SI unit,
also available as ``instance.value``. Temperature uses kelvin and the
instrument's existing temperature sensor configuration.
General also provides range/autorange, integration in power-line cycles,
averaging enable/count/type, AC bandwidth, frequency/period gate time,
trigger delay and communication timeout. Controls follow the function:
temperature has no manual range, frequency/period use gate time rather
than NPLC or averaging, and AC NPLC applies only at 300 Hz bandwidth.
Numeric expressions are evaluated at configuration time; inactive controls
are ignored. Averaging can combine conversions into the one returned
reading. Immediate triggering and a sample count of one are automatic.
Each execution also publishes the applicable numeric **Range**,
**Autorange**, **NPLC**, **Filter Enabled**, **Filter Count**, **Filter Type**,
**AC Bandwidth**, **Gate Time**, **Trigger Delay** and **Timeout** outputs.
Settings are queried after the measurement, so Range follows autoranging.
Filter Type is 0 for repeating and 1 for moving; booleans are 0 or 1.
Timeout is the local transport timeout. Settings and readings can be used
by scan data collection, Plot Points, Save and downstream expressions.
The plugin owns its session: startup connects/configures, every execution
requests a fresh conversion, and cleanup aborts/closes the connection.
Successful cleanup retains the last reading and settings; a failed read
clears them. Reconfigure may defer setup. The 2700 is used only as a
multimeter with its currently selected input terminals. No reset, relay,
route, scan-list or channel-list commands are sent. Put it in ordinary
DMM mode before use; scanner/switching support is deferred. **About**
documents these controls and outputs.
Attributes:
driver (str):
``keithley_2000`` (default) or ``keithley_2700``.
resource (str):
GPIB resource, initially ``GPIB0::16::INSTR``.
function (str):
SCPI measurement function, initially ``VOLT:DC``.
autorange (bool):
Automatic range selection, initially True.
input_range (float | str):
Fixed range in the function's SI unit, or expression.
nplc (float | str):
Integration in PLC, 0.01..10; initially 1.0.
filter_enabled (bool):
Whether to average conversions, initially False.
filter_count (int):
Averaging count, 1..100; initially 10.
filter_type (str):
``REP`` (repeating, default) or ``MOV`` (moving).
ac_bandwidth (float):
AC lower bandwidth in Hz: 3, 30 or 300 (default).
gate_time (float | str):
Frequency/period gate time in seconds, 0.01..1; initially 0.1.
trigger_delay (float | str):
Conversion delay in seconds, initially zero.
timeout (float | str):
Communication timeout in seconds, initially 30.
value (float):
Last reading, initially NaN; not persisted.
settings_readback (dict[str, float]):
Queried numeric settings from the last execution, not persisted.
instance_name (str):
Inherited Python identifier in scripts and QtConsole.
comment (str):
Inherited optional step note.
sequence_engine (SequenceEngine | None):
Inherited owning engine and live expression namespace.
has_lifecycle (bool):
True: sequence setup and cleanup manage this session.
Keyword Parameters:
parent (QObject | None): Optional Qt parent object.
Examples:
With an existing sequence instance named ``read_multimeter``, edit
settings before running and inspect the last completed reading::
read_multimeter.driver = "keithley_2700"
read_multimeter.function = "FRES"
read_multimeter.nplc = "integration_cycles"
read_multimeter.value
read_multimeter.settings_readback
"""
def __init__(self, parent=None) -> None:
super().__init__(parent)
self.driver = "keithley_2000"
self.resource = "GPIB0::16::INSTR"
self.function = "VOLT:DC"
self.autorange = True
self.input_range: float | str = 10.0
self.nplc: float | str = 1.0
self.filter_enabled = False
self.filter_count = 10
self.filter_type = "REP"
self.ac_bandwidth = 300.0
self.gate_time: float | str = 0.1
self.trigger_delay: float | str = 0.0
self.timeout: float | str = 30.0
self.value = math.nan
self.settings_readback: dict[str, float] = {}
self._meter: Keithley2000 | None = None
self._connected_driver: str | None = None
self._configured_function: str | None = None
self._apply_initial_config()
@property
def name(self) -> str:
return "Read Multimeter"
@property
def has_lifecycle(self) -> bool:
return True
def _active_numeric_settings(self) -> list[str]:
function = DmmFunction(self.function)
names = ["trigger_delay", "timeout"]
if function in Keithley2000.RANGED_FUNCTIONS and not self.autorange:
names.append("input_range")
if Keithley2000.uses_nplc(function, self.ac_bandwidth):
names.append("nplc")
if function in Keithley2000.GATED_FUNCTIONS:
names.append("gate_time")
return names
[docs]
def validate(self) -> bool:
"""Check model, function and applicable literals/expressions without I/O."""
if self.driver not in DMM_DRIVERS:
raise ValidationError("Select a Keithley 2000 or 2700 driver.")
try:
function = DmmFunction(self.function)
except ValueError as exc:
raise ValidationError("Select a supported multimeter function.") from exc
if not self.resource.strip():
raise ValidationError("A multimeter GPIB resource is required.")
if function in Keithley2000.AC_FUNCTIONS and self.ac_bandwidth not in (3.0, 30.0, 300.0):
raise ValidationError("AC bandwidth must be 3, 30 or 300 Hz.")
if function not in Keithley2000.GATED_FUNCTIONS:
if self.filter_type not in ("REP", "MOV") or not 1 <= self.filter_count <= 100:
raise ValidationError("Filter requires REP or MOV and a count of 1..100.")
for name in self._active_numeric_settings():
value = getattr(self, name)
if isinstance(value, str):
validate_expression(
value, self.engine_namespace, label=f"{self.instance_name} {name}"
)
else:
self._check_numeric(name, float(value))
return True
@staticmethod
def _check_numeric(name: str, value: float) -> None:
lower, upper = _NUMERIC_SETTINGS[name]
if not math.isfinite(value) or not lower <= value <= upper:
raise ValidationError(f"Multimeter {name} must be between {lower:g} and {upper:g}.")
def _clear_reading(self) -> None:
self.value = math.nan
self.settings_readback = {}
[docs]
def connect(self) -> None:
"""Connect and identify an owned meter, releasing partial connections."""
self.validate()
transport = GpibTransport.from_resource_string(self.resource, timeout=30.0)
try:
self._meter = DMM_DRIVERS[self.driver](transport)
self._meter.connect()
self._meter.confirm_identity()
except Exception:
self._meter = None
transport.close()
raise
self._connected_driver = self.driver
[docs]
def execute(self) -> None:
"""Publish one fresh reading and its applicable numeric setting readbacks."""
self._clear_reading()
if self._meter is None or self._connected_driver != self.driver:
raise RuntimeError("The selected multimeter is not connected.")
if self._configured_function != self.function:
raise RuntimeError("Reconfigure the multimeter for this measurement function.")
value, settings = self._meter.read_single_snapshot(DmmFunction(self.function))
self.value = float(value)
self.settings_readback = settings
[docs]
def disconnect(self) -> None:
"""Abort and close this instance's session, retaining successful outputs."""
meter, self._meter = self._meter, None
self._connected_driver = self._configured_function = None
if meter is not None:
try:
meter.abort()
finally:
meter.disconnect()
def _setting_outputs(self) -> dict[str, tuple[str, str]]:
function = DmmFunction(self.function)
outputs = {"trigger_delay": ("Trigger Delay", "s"), "timeout": ("Timeout", "s")}
if function in Keithley2000.RANGED_FUNCTIONS:
outputs.update(range=("Range", DMM_CHANNELS[function][1]), autorange=("Autorange", ""))
if Keithley2000.uses_nplc(function, self.ac_bandwidth):
outputs["nplc"] = ("NPLC", "PLC")
if function in Keithley2000.GATED_FUNCTIONS:
outputs["gate_time"] = ("Gate Time", "s")
else:
outputs.update(
filter_enabled=("Filter Enabled", ""),
filter_count=("Filter Count", ""),
filter_type=("Filter Type", ""),
)
if function in Keithley2000.AC_FUNCTIONS:
outputs["ac_bandwidth"] = ("AC Bandwidth", "Hz")
return outputs
[docs]
def declared_outputs(self) -> PluginOutputs:
"""Declare configured outputs and their physical units together."""
var = self.instance_name
label, unit = DMM_CHANNELS[DmmFunction(self.function)]
outputs = PluginOutputs(values={f"{var}:{label}": ScalarOutput(f"{var}.value", unit)})
for name, (label, unit) in self._setting_outputs().items():
outputs.values[f"{var}:{label}"] = ScalarOutput(
f"{var}.settings_readback.get({name!r}, float('nan'))", unit
)
return outputs
[docs]
def to_json(self) -> dict[str, Any]:
"""Persist editable settings, excluding sessions and acquired data."""
data = super().to_json()
data.update({name: getattr(self, name) for name in _SAVED_SETTINGS})
return data
def _restore_from_json(self, data: dict[str, Any]) -> None:
"""Restore editable settings while retaining defaults for absent fields."""
super()._restore_from_json(data)
for name in _SAVED_SETTINGS:
if name in data:
setattr(self, name, data[name])