6221/2182A trace ================ Choose **6221/2182 IV Trace** from the Keithleys catalogue for a current-driven I-V curve with a Keithley 6221 and primary 2182A nanovoltmeter. The scan generator supplies a current list in amperes; Trigger Link handshaking coordinates the source and buffered readings. This is a complete trace acquisition. Use :doc:`keithley_point_commands` when the source must hold each DC level while child steps run. Connections and settings ------------------------ On **Settings > Primary 6221 / 2182A**, select the 6221 GPIB resource and choose how to reach the 2182A: through the 6221 serial relay, or through its own GPIB resource. Serial relay carries commands; Trigger Link still carries the measurement handshake. .. figure:: _static/screenshots/keithley-6221-settings.png :alt: Dark 6221/2182A Settings page showing connections, compliance, current range, integration, filters and trigger lines. :width: 100% Primary source and meter configuration. The secondary meter is enabled for this illustrated configuration, but connects only when the sequence runs. Choose fixed-voltage compliance or the per-point limit derived from ``abs(current) * compliance_resistance``. Configure the settling delay and current-range policy: **Best** selects a range for the whole list, **Auto** permits per-point range changes, and a fixed range holds the selected value. The meter controls cover NPLC integration, voltage range, digits, trigger delay, autozero, line synchronisation, digital/analogue filters and relative mode. Runtime expressions are evaluated by the acquisition configuration, not by the scan preview. The source-to-meter and meter-to-source trigger lines must match the wiring. Defaults are line 2 from the 6221 to the 2182A and line 1 for the return handshake. See :doc:`keithley_triggers` before changing them. Optional second nanovoltmeter ----------------------------- **Secondary nanovoltmeter** enables an independently connected Keithley 182 or 2182A. Set its GPIB resource, column prefix, supported integration/range/ filter settings and trigger routing. Controls follow the selected driver. .. figure:: _static/screenshots/keithley-6221-detail.png :alt: Dark Secondary nanovoltmeter page with independent meter settings and parallel or daisy-chain triggering. :width: 100% Secondary-meter configuration for the ordinary 6221/2182A trace. In **parallel** routing, both meters receive each source trigger; the primary returns the handshake. The secondary must finish before the next trigger. The configuration warns when estimated integration/filter time is too long. In **daisy-chain** routing, the primary triggers the secondary, and the secondary's completion advances the source. This waits for both conversions. See the routing guide for the separate trigger lines. Scan and results ---------------- **Scan** selects the current generator. The ``IV`` trace has programmed current ``x`` (A), ``V`` (V), ``R`` (ohms) and ``P`` (W). Ordinary-mode resistance is ``V / I`` and is NaN at zero current; power is ``I * V``. Enabling the second meter adds prefixed voltage, response and power columns to the same table. Plot Trace and Save consume this complete result; see :doc:`data_workflows`. **Differential mode** alternates positive and negative **Delta current** offsets around the nominal scan currents. It needs at least three points and a positive delta current. Returned ``x`` values remain nominal; neighbouring readings supply reduced voltage and differential ``R`` or ``G`` (siemens), depending on the resistance/conductance choice. Power uses delta current times the reduced voltage change. Endpoint responses use the nearest interior reduction. This is separate from :doc:`keithley_pulse_delta`. Acquisition and cleanup ----------------------- The plugin programs the current list and retrieves buffered meter readings after each sweep. It owns the source, primary-meter connection or serial relay, and any enabled secondary-meter session. Disconnect attempts to disable the source output and close those sessions. Do not use the buffered sweep as a substitute for a held-DC point scan. For the Python API and console attributes, find ``Keithley6221_2182APlugin`` in the :doc:`plugin API reference `. Return to :doc:`keithley_meters` to compare the available acquisition modes.