Console and script reference

This reference is for experienced users inspecting results or developing Python measurement scripts. The console on Script Editor shares the sequence engine’s live namespace. Examples below assume the named steps are present; use the downloadable sequences in Usage and Collecting, fitting and saving data. For implementing an external plugin, follow Writing sequence plugins.

Inspecting the active sequence

The application exposes a live sequence view, not a snapshot:

sequence.steps              # nested plugin/(plugin, children) descriptors
sequence.top_level_plugins  # root instances only
sequence.plugins            # instances at every depth, in tree order
[(p.instance_name, p.name) for p in sequence.plugins]

Instances are also bound by their names, for example iv_trace or curve_fit. The first step may be Details or a command; do not assume sequence.plugins[0] is an instrument or has a scan generator.

Runtime names and expressions

The namespace contains np/numpy, exported NumPy functions, log, plugin instance variables and the _traces/_values catalogues. A catalogue maps a selectable identity to an expression rather than containing the measurement itself. The sequence view is provided by the application; a standalone SequenceEngine does not automatically have an editor tree.

_traces
_values
iv_trace.data['Dummy'].df.head()
iv_trace.log.info("Inspecting the current trace")

Use actual instance names from the active sequence. A result is available only after its producing step has completed, and a failed measurement/fit may expose empty data or NaNs. Plot tables and scalar outputs are described in Trace data and output catalogues.

Editing settings and refreshing controls

Make configuration edits while execution is idle. A direct attribute edit can affect the next execution without immediately rebuilding its cached widgets. Use plugin.refresh() after edits, plugin.full_refresh() for that container and its descendants, or sequence.full_refresh() for the tree. These refreshes synchronise configuration UI and the engine’s output catalogues, including configured trace channels and units. They do not acquire new data or connect to or configure instruments.

refresh() updates existing configuration pages only where their widgets provide a refresh hook; it does not rebuild pages or synchronise every raw editor automatically. The catalogues are rebuilt before those widgets refresh. For output identities, units and configured channel schemas, see Plugin output declarations. Their legacy dictionary views remain available in the console, but catalogue entries can refer to data that have not been acquired.

For the Function Scan Generator in the basic I-V example:

iv_trace.scan.amplitude = "5e-6"
iv_trace.scan.num_points = 201
iv_trace.refresh()

scan is an alias for scan_generator only when the plugin has one. Generator properties differ: a function generator has waveform parameters, a list generator has stages, and a stepped generator has its own stage definition. Inspect the actual generator rather than assigning generic start/end fields to every type. For the Counter list in the workflow:

scan.scan.stages = [(0.0, True), (1.0, True), (2.0, True)]
scan.refresh()

A state sweep uses sweep_generator instead. Preserve the boolean measurement flag when editing stages. If settings change reported output names, refresh data catalogues through the attached engine as described in Writing sequence plugins.

Running code and lifecycle readiness

Console method calls are real actions on shared objects. They can move hardware or write files. For a repeatable check with normal setup and cleanup, use Run Selection…; for a modified full script, use Generate Code and run the edited script on Script Editor. See Sequence execution reference.

After generated cleanup, a lifecycle-managed instance is no longer ready for arbitrary action calls. Reconnecting begins a new session and may reset per-run state. In particular, finish an incremental Save before cleanup if it must retain the original destination; see Collecting, fitting and saving data.

Selecting a plot display

The application injects a display-selection helper:

select_display_plugin('polar')

Selecting the current display preserves it. A different selection is dispatched to the GUI and can clear incompatible plotted state; acquired tables remain available. An unavailable requested display raises an error. Generated scripts use this helper before plugin setup to select the sequence’s preferred display. Use Plot displays and controls for coordinate forms and user controls, and Plot display communications contract for direct producer requests.

Temperature controller scripting

The shared controller service is separate from the sequence engine. Obtain it explicitly; engine is not an automatically injected controller variable:

from stoner_measurement.temperature_control.engine import TemperatureControllerEngine
from stoner_measurement.temperature_control import LoopRef, ChannelRef

temperature = TemperatureControllerEngine.instance()
temperature.set_setpoint(1, 300.0)  # primary controller, native loop 1
temperature.set_setpoint(LoopRef('secondary', 1), 275.0)
temperature.set_input_channel(
    LoopRef('secondary', 1), ChannelRef('secondary', 'B')
)
state = temperature.get_engine_state()
state.all_readings[ChannelRef('secondary', 'A')].value
state.loop_values('setpoints')[LoopRef('secondary', 1)]

These examples require the corresponding connected controllers and supported channels. temperature.controller('secondary') exposes controller-local settings. Legacy integer-keyed state dictionaries and state.readings remain primary views; state.secondary contains the secondary snapshot. cache_ages and controller_statuses distinguish freshness and status.

For a Set Temperature step and a temperature-monitor instance:

set_temperature.controller_id = 'secondary'
set_temperature.control_loop = 1
set_temperature.setpoint_expr = '275'
temperature_monitor.temperature({'controller_id': 'secondary', 'channel': 'A'})

Use the actual step names and refresh their UI after edits. Do not close the shared service as cleanup for a plugin that merely borrows it. The connection, stability and migration controls are covered in Temperature controllers.