Temperature controllers

This is the user and rig-administrator reference for the Temperature Control panel, loop selections and shared stability settings. For qualified Python calls see Temperature controller scripting; for plugin contracts see Writing sequence plugins. See Controllers for the common engine, panel and main-window status-bar workflow.

A rig can use one temperature controller or a primary controller plus an optional secondary controller. The secondary controller is disabled by default. Existing single-controller settings and sequences continue to select primary.

Connecting instruments

Open Connection in the Temperature Control panel. Use the Primary and Secondary sub-tabs to select each controller’s driver, transport and address. For a rig with a second controller, select Enable secondary controller on the Secondary sub-tab to make its form available. Each controller has its own Connect and Disconnect buttons.

Temperature Control Connection page with the Secondary tab selected and its connection form enabled.

The Secondary connection form, shown with a built-in simulated controller.

The main-window Temp indicator’s Connect and Disconnect shortcuts address the primary slot. Use the panel’s secondary controls to change the secondary connection; disconnecting primary does not disconnect secondary.

The shared polling rate and Save Settings to YAML appear below both sub-tabs. Saving records both connection forms, including settings for an instrument that has yet to connect. Disabling secondary disconnects it while retaining its saved settings. A polling rate of zero disables automatic polling. Immediate reads still work.

The service rejects using the same physical connection in both slots. Different instances of the simulated temperature controller can be used for testing.

Loops, sensors and settings

The Control page has Primary and Secondary tabs, each showing that controller’s loops side by side. The Secondary tab is hidden while secondary is disabled. Loops are labelled by controller and native loop number. A loop’s control-sensor list contains only inputs supported by that loop on its own instrument. Identical names such as channel A or loop 1 remain distinct on the two controllers. Disconnecting or replacing one controller does not renumber the other controller’s loops.

Zone-table selectors include the owning controller. A zone table is read when the Zone Table page is first opened or its selected loop changes, rather than during connection. Use Read for an explicit refresh.

Input Settings has Primary and Secondary sub-tabs, matching the Control page. Each controller’s input settings and calibration-curve names load only when its sub-tab is first accessed. Returning to a loaded page preserves edits; use Read All to refresh it. Reconnecting a controller invalidates its settings, without reloading the other controller’s settings or discarding its edits.

Unread settings are greyed out, remain greyed out while loading, and become editable after a successful read. Failed reads leave the controls and Apply or Write buttons disabled; Read and Read All remain available to retry. Cryogen controls appear for each instrument that supports them. Heater ranges and other optional features belong to the selected instrument.

Panel connections, disconnections, reads, Apply and Write All run in a background temperature-operation queue. Other panels remain responsive while a controller or a large zone table is loading. Requests are serialised with polling and other operations on the same session. Repeated Read clicks do not build up duplicate work, and results for a previous connection or selection are discarded.

Controls stay greyed out until the read finishes successfully. A requested disconnect waits for any native instrument call already running; it does not interrupt that call. Closing the panel hides it and allows pending work to finish. Application shutdown drains running work before releasing the controller sessions. Console and plugin calls to the engine retain their synchronous behaviour.

For an Oxford ITC503, Input Settings offers an NI-DAQmx custom-scale selector for each sensor A, B and C, in either controller slot. The scale must be a Table scale whose two columns contain temperatures in kelvin. Linear, polynomial and voltage/resistance-to-temperature scales cannot be imported here. Use Refresh NI Scale List to discover scales saved in NI MAX on this computer. Discovery and importing require the optional nidaqmx package and NI-DAQmx runtime.

Select ITC503 → Corrected when the NI table’s pre-scaled column contains ITC503 readings and its scaled column contains corrected temperatures. Select Corrected → ITC503 for the opposite definition. None disables conversion for that sensor. Existing YAML lookup retains the previous shared lookup from instruments/oxford_itc503.yaml; existing configurations keep using it until you explicitly change their selections.

Apply validates all selected tables, activates the software conversion and saves the scale names, directions and table values under each controller’s itc503_calibration entry in temperature_controller.yaml. The imported snapshot works after restarting without NI-DAQmx. Applying an already saved scale uses its saved table. To import a changed NI table with the same name, check Reload selected tables from NI-DAQmx on Apply before applying. NI scales themselves are never modified.

Imported tables use piecewise linear interpolation. Both temperature columns must contain at least two finite, non-negative, strictly increasing values; outside the table’s range the temperature is left unchanged. Sensor readings use their individual scales. Setpoints and PID-zone temperatures use the scale of the heater sensor selected on the ITC503, including inverse conversion when writing a corrected setpoint. Applying a scale does not rewrite a hardware setpoint; loop settings are refreshed and cached zone-table values are discarded.

For a Lake Shore 340, Input Settings provides native controls for each A/B input, including input enable, sensor type, excitation, voltage range, compensation, filtering and calibration curve. It works in either controller slot. Use Read All to load the hardware settings before writing.

Choose a sensor preset to let the instrument select its associated units, coefficient, excitation and range. Choose Special to edit those values directly. Compensation retains the separate Off, On and Paused states. Write All sends only changed settings and reads them back, including any defaults selected by a new sensor preset. After a failed write or readback, use Read All before retrying. Polling does not overwrite these edits.

Dark Temperature Control page with Primary and Secondary tabs and two primary loop control groups.

Loop controls with the built-in simulator; the secondary controller has its own labelled tab.

The Stability page has one shared table for the whole rig. Each tolerance and rate sensor can come from either controller: A is primary channel A and A.2 is secondary channel A. An empty selection uses the first available sensor on the rig, primary first. An explicit missing sensor, invalid reading, failed poll or stale state cannot satisfy a stability wait. A controller’s failure also invalidates any peer loop whose criteria depend on its readings. Both controllers are read before stability is evaluated.

Rows are evaluated in their displayed order against each loop’s target setpoint. The first row whose upper temperature limit includes the target wins, even when another row also matches. Use Move Up and Move Down to choose priority; rows are never automatically sorted. If no row matches, the last row applies. Changing the selected band or sensor starts a fresh stability window.

Dark Stability page showing the shared ordered tolerance and rate-sensor table and row-priority controls.

One ordered stability table for the whole temperature rig.

The chart combines temperature, setpoint, heater and supported gas-flow traces. Secondary labels use the suffix .2, such as T_A.2 and SP_1.2. Internal trace keys retain their previous names to preserve saved styles and visibility. Drag the divider between the plot and legend to resize them.

There is one dT/dt trace. It uses the engine’s rate for the selected rate sensor in the active stability band. When loops use different bands, the earliest active table row takes priority; ties use primary before secondary, then increasing loop number. Hover over its legend label to see the chosen sensor. Sensor changes clear the rate history to avoid joining unrelated measurements. Missing or invalid selected data clears the trace and displays Unavailable.

Sequences and console use

Set Temperature, Temperature Controller Scan and Temperature Controller Sweep use a controller-labelled loop selector. Saved selections remain visible as unavailable while disconnected. A missing secondary target never selects primary instead. Scan/sweep limits come from the chosen controller.

Sensor and monitor-loop lists have catalogue pickers. Editable lists accept A, secondary:A for sensors and 1, secondary:1 for loops. Unqualified entries mean primary. An empty sensor list includes all available sensors. The monitor leaves the shared controller connections running when it stops.

For controller-qualified Python calls and state inspection, see Temperature controller scripting. Legacy unqualified loop numbers and sensor names continue to mean primary; use qualified references for the secondary device.

Make Safe attempts output shutdown on every advertised loop of both connected controllers, including after a failure on the first. Stopping the temperature engine attempts to disconnect both. These operations report failures; they do not substitute a different target or claim successful writes after an I/O error.

Configuration compatibility

The machine YAML uses schema_version: 3 with separate connection settings in controllers.primary and controllers.secondary, and one top-level stability table. Stability sensor keys retain secondary:A internally; .2 is their compact display suffix.

Older per-controller stability settings migrate using the primary table as the shared table. An explicit version 3 top-level table takes precedence. Existing single-controller settings and timestamped backups remain supported. Review the shared table when migrating a rig that previously used different secondary criteria.

Old sequence loop numbers and sensor strings mean primary. New command and state plugin JSON includes controller_id. Mixed monitor and sensor selections store structured references. Sequence JSON and saved-data sequence metadata retain those identities. Two-controller sequences require a release supporting this schema; older releases cannot be expected to read them.

Timer-driven polling is dispatched to a controller worker so device I/O does not block the GUI event loop. Shared-state reads still need controller data; a slow or failed instrument can delay fresh readings and invalidate dependent stability criteria. Transport timing, actual sensor assignments and output shutdown must be checked on the measurement rig before relying on a new hardware configuration.