Keithley connections and triggering¶
Choose the acquisition mode before wiring triggers. The four trace plugins use different coordination schemes; a GPIB address or serial relay selects a command path, while Trigger Link or a reference cable carries timing signals. Return to Keithley meters for the measurement choices.
6221/2182A buffered I-V¶
The ordinary 6221/2182A trace trace programs a 6221 current list. After settling at each point, the source triggers the primary 2182A; its meter-complete response permits the next source step. Both buffered meter readings and source progression must complete before the trace is returned.
The primary settings expose the two Trigger Link lines:
6221 to 2182A: source output to primary-meter external trigger input (default line 2).
2182A to 6221: meter-complete output to source input (default line 1).
Selecting Via 6221 serial port sends meter commands through the source’s RS-232 relay. It still needs the Trigger Link connection for this handshake. Selecting Direct GPIB gives the primary its own GPIB session; it does not remove the hardware-trigger requirement.
With a second nanovoltmeter enabled, choose one topology on Secondary nanovoltmeter:
Parallel: the source triggers both meters; the primary alone returns completion to the source. The secondary must finish before the next source trigger. Increase settling/primary conversion time or shorten secondary integration/filtering when the timing warning indicates an overrun.
Daisy-chain: source triggers primary, primary completion triggers secondary, and secondary completion advances source. Configure the primary-to-secondary and secondary-to-source lines to match the cable. Source progression then waits for both meters.
Routing is configured with the secondary meter’s conversion settings.¶
2400 trigger model¶
The 2400 buffered trace, point scan and set command expose the supported trigger-model controls on the nested Advanced page.
Source |
Use |
|---|---|
Immediate |
Proceed through the instrument’s configured acquisition without an external start signal. |
Bus |
Let the plugin issue the bus trigger when starting the acquisition. |
External / Trigger Link |
Wait for the corresponding external routing; configure the input line and provide the expected signal. |
Timer |
Use the configured instrument timer interval. |
The output-enable checkbox and output-line selector are separate from the choice of incoming trigger source. Arm count, trigger-count override and delays also affect completion. An unprovided external trigger can leave an acquisition waiting until its timeout.
When the buffered 2400 trace enables its secondary nanovoltmeter, it configures a dedicated SMU/meter Trigger Link handshake. The SMU’s output line triggers the meter and the meter’s completion returns on the configured input line. This acquisition chooses its own compatible counts and arm behaviour rather than combining arbitrary manual trigger overrides. Meter integration and filtering must fit the intended cadence. See 2400 trace for settings and result columns.
6221 with lock-ins¶
The 6221/lock-in trace trace uses a 6221 sine-wave phase marker on its selected Trigger Link line as the external reference for the lock-ins. This is an AC phase/frequency reference, rather than a buffered point-complete handshake. The plugin changes the chosen source parameter, allows the configured time-constant/cooldown settling and reads the selected outputs. Choose the harmonic, phase and input configuration for each lock-in.
Native Pulse Delta¶
6221/2182 Pulse Delta trace requires both RS-232 and Trigger Link between the 6221 and 2182A, using the native mode’s factory Trigger Link assignments. Both serial ports use 19.2 kbaud without flow control; the PC talks to the 6221 over GPIB. The source manages the pulse/meter handshake and integration.
The ordinary I-V trace’s arbitrary secondary-meter routing does not apply to native Pulse Delta. That plugin supports no auxiliary meter. Its live timing diagram shows nominal pulse and baseline timing, with a 50/60 Hz preview assumption; it does not change the instrument’s mains configuration or verify physical cable timing.
Held-DC point scans and set commands¶
The 6221 point scan and Set 6221/2182 command directly set a DC level, wait for settling, then read any enabled meters sequentially. They do not emit output triggers or invoke LIST/Pulse Delta merely to obtain a pulse. Current remains held while nested or later steps execute, until it is changed or cleanup disables the source. The primary and secondary meter sessions are optional.
The 2400 point scan and Set 2400 retain their instrument trigger-model options, including Trigger Link output. They apply one source level at a time, rather than programming a buffered source sweep. See Keithley point scans and set commands for per-point outputs and lifecycle behaviour.
For any physical rig, check the cable routing, signal compatibility, integration/filter settings and source limits against the selected instruments. Configuration screenshots illustrate the software controls; they do not establish bench validation of the measurement timing.