Keithley meters¶
Keithley measurement steps have their own instrument connections and settings. Choose a trace for a complete acquired curve, a point scan to run child steps at each source level, or a set command to hold one bias for later steps. These steps do not borrow the temperature, magnet or motor engines described in Controllers.
Choose a measurement¶
Guide |
Measurement |
Returned trace |
|---|---|---|
Current list from a 6221; synchronised buffered voltage from an 2182A, with an optional second nanovoltmeter |
|
|
6221 sine amplitude, DC offset or frequency scanned with one or more SR830 or SIGNAL RECOVERY 7265 lock-ins |
|
|
Buffered current- or voltage-source LIST sweep on a 2400 SMU, with an optional second nanovoltmeter |
|
|
Native 6221/2182A pulse acquisition with one or two baseline readings |
|
The ordinary 6221/2182A, current-source 2400 and native Pulse Delta traces also offer alternating delta-current acquisition for differential resistance or conductance. This optional setting changes the response columns; it does not make the ordinary I-V trace use the native Pulse Delta timing model. Each guide explains the mode and its returned columns.
Use Keithley point scans and set commands for 2400 Point by Point Scan, 6221/2182 Point by Point Scan, Set 2400 and Set 6221/2182. See Keithley connections and triggering for trigger routing and the distinction between buffered handshakes, AC reference signals, native pulses and held DC.
Single electrometer readings¶
Choose Keithleys → Read Electrometer to take one fresh reading from a Keithley 6485 picoammeter or 6514/6517 electrometer. On General, choose the model, GPIB resource, measurement function, range/autorange, integration, averaging, zero check, trigger delay and timeout. The 6485 reads current; the 6514/6517 can also read voltage, resistance or charge. Numeric expressions are evaluated at configuration time.
Each execution publishes one scalar output named Current, Voltage,
Resistance or Charge, with its SI unit. Select that output in an outer
scan’s data collection or use it in Plot Points, Save and later expressions.
For an instance named read_electrometer, the most recent reading is also
available as read_electrometer.value. Failed reads clear it to NaN.
Startup opens and configures the command’s owned connection; cleanup closes
it. Every execution takes a new measurement using immediate triggering,
without external trigger wiring or buffer acquisition. Averaging may combine
several conversions into that one result. See the
ReadElectrometerCommand
API documentation for script attributes and console examples.
Single multimeter readings¶
For a single 2000/2700 multimeter reading, choose Keithleys → Read Multimeter. On General, select the model, GPIB resource and function. Controls include range/autorange, NPLC, averaging, AC bandwidth, frequency/period gate time, trigger delay and timeout, with unavailable settings disabled. AC NPLC applies at 300 Hz bandwidth; frequency/period use gate time. Temperature uses kelvin and the instrument’s existing sensor configuration.
Each execution publishes the reading and applicable numeric setting readbacks
(including the actual range after autoranging). For an instance named
read_multimeter, inspect read_multimeter.value and
read_multimeter.settings_readback. Settings can be selected as scalar
outputs alongside the measurement. Filter Type is 0 for repeating and 1 for
moving; boolean settings use 0/1. Timeout is the local transport setting.
The 2700 is used as a multimeter on its currently selected input terminals. No relay, routing, scan-list, channel-list or reset commands are sent; configure it for ordinary DMM operation beforehand. Switching support is deferred.
Common workflow¶
Add the required step from the Keithleys catalogue.
Set the instrument connections, ranges, compliance, integration and timing on Settings for traces/point scans, or General for set commands. Trace and point-scan plugins choose source values on Scan; set commands take a Source value instead.
For traces, add Plot Trace and Save steps to use the acquired result. For point scans, put the required measurement steps inside the scan and select the scalar outputs to collect. See Collecting, fitting and saving data.
Run the sequence. Startup connects and configures the owning instances; cleanup releases their sessions. Ordinary DC/AC steps may retain output between executions; Pulse Delta aborts to zero/off after each acquisition.
Use one owning plugin instance per physical source in a sequence. Reconfigure can defer an instance’s connection/configuration until its execution point; see Reconnecting and applying settings at a chosen step. Screenshots in these guides show the real dark-mode configuration pages with example addresses; they do not establish that any physical instruments were connected.